INDOEX HARMONY WORKSHOP
– DATA SUMMARY SHEET
A. Clarke and S. Howell, University
of Hawaii, 808-956-6215
[tclarke@soest.hawaii.edu]
C-130 – INSTRUMENTATION
(onboard behind sampling inlet)
LASER OPTICAL PARTICLE COUNTER
[OPC]
-
256 size intervals @ nominal 0.1 < Dp < 7.0m
m; 1-3min resolution
-
Thermal analysis – sizes at approx. 30, 150, 300C(Refractory).
Volatility at 150C suggests sulfuric acid or low temp volatile species
present while residue at 300C suggest refractory components including soot.
-
Calibration based upon ammonium sulfate non-absorbing
aerosol needs correction for other refractive indices – including 300C
sootlike residuals.
-
Reduced counting efficiency in 0.1-0.2m
range – needs correction
-
C-130 Inlet losses expected for sizes increasing above
about 3m m
-
Operated "dry" – usually 20-40%RH by using 50/50 dilution
flow
-
Flow uncertainties estimated at +/- 10%
DIFFERENTIAL MOBILITY ANALYZER [DMA]
-
40 size intervals; nominal 0.006 < Dp < 0.25m
m; 1-3 min. resolution
-
Thermal analysis – sizes at approx. 30, 150, 300C.
-
Calibration intrinsic but checked with latex spheres
-
Some losses expected at smaller sizes below about 0.01m
m
-
Some count errors expected above about 0.2m
m
-
Operated "dry" usually at less than 20%RH using dried
sheath air flow
-
Flow uncertainties estimated at +/- 10%
Note: DMA and OPC number at overlap
near 0.2m usually differ only by about 5-15%
CONDENSATION NUCLEUS COUNTER
[CNambient, CNhot-300deg.C and UltrafineUCN]
-
CN and CNhot are number concentrations for nominal Dp>13nm,
1Hz
-
CNhot/CN ratio can be used as indicator of polluted vs.
clean air
-
Ultrafine UCN are number concentrations for nominal Dp>3nm
-
Flow uncertainties nominally +/-5%
-
Elevated counts can occur in cloudy regions due to droplet
shatter on inlet
Note: Total of integral OPC + DMA
number concentrations should agree with UCN within flow uncertainties –
and generally do within about 10-15%